á1051ñCLEANING GLASS APPARATUS
Change to read:
Success in conducting many Pharmacopeial assays and tests depends upon the utmost cleanliness of the glassware apparatus used.
![]() ![]() One effective method used in the past for cleaning glassware is the application of hot nitric acid.Asecond traditional method for removing organic matter that does not require heat is the use of a chromic acidsulfuric acid mixture.However,the chromic acid wash is not recommended because of the hazardous and toxic nature of the material.
Several safer alternatives,including the use of cleansing agents,such as trisodium phosphate and synthetic detergents,have proven highly useful,but require prolonged rinsing.
![]() ![]() For optical measurements,special care is required for cleaning containers,but the use of both chromic acid and highly alkaline solutions should be avoided.
Effective removal of organic matter is very important for testing pharmaceutical waters in accordance with the general test chapter Total Organic Carbon á643ñ.It has been demonstrated that an alkaline detergent with potassium hydroxide as the primary ingredient*leaves the least amount of organic matter residuals.Heating in a muffle furnace produces comparable results and is the least labor-intensive procedure;however,it requires specialized equipment.
In all cases,it is important to verify that the cleaning procedure is appropriate for the particular test or assay being undertaken.This can be accomplished via blank runs,scientific judgments,residuals data from cleansing agent and detergent manufacturers,or other controls.Specifically,special care is required for cleaning containers for optical measurement applications;the use of highly alkaline and the no longer recommended chromic acid solutions should be avoided.
![]() ![]()
*