Medical Air
»Medical Air is a natural or synthetic mixture of gases consisting largely of nitrogen and oxygen.It contains not less than 19.5percent and not more than 23.5percent,by volume,of O2.
Packaging and storage
Preserve in cylinders or in a low pressure collecting tank.Containers used for Medical Air are not to be treated with any toxic,sleep-inducing,or narcosis-producing compounds,and are not to be treated with any compound that would be irritating to the respiratory tract when the Medical Air is used.
NOTEReduce the container pressure by means of a regulator.Measure the gases with a gas volume meter downstream from the detector tube in order to minimize contamination or change of the specimens.
The various detector tubes called for in the respective tests are listed under Reagent Specificationsin the section Reagents,Indicators,and Solutions.
Labeling
Where it is piped directly from the collecting tank to the point of use,label each outlet Medical Air.
Water and oil
Support 1container in an inverted position (with the valve at the bottom)for 5minutes.Cautiously open the valve slightly,maintaining the container in an inverted position.Vent the gas with a barely audible flow against a stainless steel mirror for a few seconds:no liquid is discernible on the mirror.
Odor
Carefully open the container valve to produce a moderate flow of gas.Do not direct the gas stream toward the face,but deflect a portion of the stream toward the nose:no appreciable odor is discernible.
Carbon dioxide
Pass 1000±50mLthrough a carbon dioxide detector tube at the rate specified for the tube:the indicator change corresponds to not more than 0.05%.
Carbon monoxide
Pass 1000±50mLthrough a carbon monoxide detector tube at the rate specified for the tube:the indicator change corresponds to not more than 0.001%.
Limit of nitric oxide and nitrogen dioxide
Pass 550±50mLthrough a nitric oxidenitrogen dioxide detector tube at the rate specified for the tube:the indicator change corresponds to not more than 2.5ppm.
Sulfur dioxide
Pass 1050±50mLthrough a sulfur dioxide detector tube at the rate specified for the tube:the indicator change corresponds to not more than 5ppm.
Assay
Determine the oxygen concentration of Medical Air using an electrochemical cell analyzer readable to 0.1%of oxygen and calibrated with ambient air to an accuracy of ±0.2%of oxygen.[NOTEThe instrument utilizes the variations of electric current produced by the interaction of oxygen with an electrochemical cell to display the oxygen strength of a confined sample or an in-line flow of the gas.This current generates a signal proportional to the oxygen concentration,which is displayed on a meter.]
Auxiliary Information
Staff Liaison:Kahkashan Zaidi,Ph.D.,Senior Scientific Associate
Expert Committee:(AER)Aerosols
USP28NF23Page 54
Pharmacopeial Forum:Volume No.28(4)Page 1065
Phone Number:1-301-816-8269
|