PROPELLANTS
CautionHydrocarbon propellants are highly flammable and explosive.Observe precautions and perform sampling and analytical operations in a well-ventilated fume hood.
General Sampling Procedure
This procedure is used to obtain test specimens for those propellants that occur as gases at about 25
![]() ![]()
Approximate Boiling Temperature
Transfer a 100-mLspecimen to a tared,pear-shaped,100-mLcentrifuge tube containing a few boiling stones,and weigh.Suspend a thermometer in the liquid,and place the tube in a medium maintained at a temperature of 32
![]()
High-Boiling Residues,Method I
Allow 85mLof the specimen to distil as directed in the test for Approximate Boiling Temperature,and transfer the centrifuge tube containing the remaining 15mLof specimen to a medium maintained at a temperature 10
![]()
High-Boiling Residues,Method II
Prepare a cooling coil from copper tubing (about 6mm outside diameter ×about 6.1m long)to fit into a vacuum-jacketed flask.Immerse the cooling coil in a mixture of dry ice and acetone in a vacuum-jacketed flask,and connect one end of the tubing to the propellant sample cylinder.Carefully open the sample cylinder valve,flush the cooling coil with about 50mLof the propellant,and discard this portion of liquefied propellant.Continue delivering liquefied propellant from the cooling coil,and collect it in a previously chilled 1000-mLsedimentation cone until the cone is filled to the 1000-mLmark.Allow the propellant to evaporate,using a warm water bath maintained at about 40
![]() ![]()
Water Content
Proceed as directed under Water Determination á921ñ,with the following modifications:(a)Provide the closed-system titrating vessel with an opening through which passes a coarse-porosity gas dispersion tube connected to a sampling cylinder.(b)Dilute the Reagentwith anhydrous methanol to give a water equivalence factor of between 0.2and 1.0mg per mL;age this diluted solution for not less than 16hours before standardization.(c)Obtain a 100-g specimen as directed under General Sampling Procedure,and introduce the specimen into the titration vessel through the gas dispersion tube at a rate of about 100mLof gas per minute;if necessary,heat the sample cylinder gently to maintain this flow rate.
Other Determinations
For those aerosols that use propellants,perform the tests specified in the individual NFpropellant monographs.
|